Introduction

The ingress of hydrogen inside structural metallic materials in engineering parts in service leads to a degradation of their mechanical properties and their premature catastrophic failures1–4. Hydrogen that freely diffuses through the material under ambient conditions5 can interact with crystalline defects and contributes to the deterioration of the mechanical properties3,5–7. A strategy to mitigate the deleterious influence of hydrogen is to design alloys with a high number density of trapping sites to limit the deleterious influence of H on moving dislocations5,6,8. Traps can even be irreversible traps, i.e. H is unable to re-enter the lattice under service conditions, owing to the high binding energy with H9,10. Trapped hydrogen has even been reported to potentially increase the resistance to hydrogen embrittlement of some materials1,2,5. In order to guide the design of hydrogen-resistant materials, it is necessary to study the details of the structure and composition of sites that can trap diffusible hydrogen, which are mostly defects such as stacking faults, dislocations and phase and grain boundaries1,5. Very few techniques have the combination of high spatial resolution and compositional sensitivity.

Atom probe tomography (APT) is a time-of-flight mass spectroscopy technique, which maps the spatial distribution of specific chemical species within a three-dimensional (3D) volume with sub-nanometre resolution11,12. In principle, APT is capable of detecting and quantifying hydrogen in three dimensions at near-atomic scale13. Yet despite some successes14–17, and decades of work from numerous research groups, hydrogen microanalysis remains very challenging1,2,13,14,17–20. There are issues associated with the influence of residual gases from the analysis chamber of atom probe, specimen preparation and transport20,21, and a strong dependence of the analytical performance on the analysis conditions22–25. Let us discuss these aspects in more details.

Studies have reported on the ionization of residual gases from the ultra-high vacuum analysis chamber of the atom probe, including hydrogen desorbed from the chamber walls, which can obscure the detection of hydrogen from the specimen20,22,25,26. To circumvent this issue, APT specimens are charged with deuterium (D or 2H) instead of hydrogen, as D has relatively lower natural abundance, i.e., 0.0156% of all hydrogen found on the earth. Residual hydrogen ionizes in the form of H+, H2+ and H3+, and is therefore detected with characteristic peaks at 1, 2 and 3 Da, respectively, and their relative amplitude depends on the intensity of the electric field at the end of the field emitter25,27. The H2+ signal interferes with the D signal, making the quantification arduous25. It is hence crucial to minimize the likelihood of detecting H2+, which, typically, involves maximizing the strength of the electrostatic field. The use of high voltage pulsing rather than laser pulsing is hence recommended so as to minimize the influence of molecular ionic species of hydrogen20,25,27.

Gemma et al. have studied the distribution of deuterium in Fe/V multi-layered thin films deposited on W18,19. They performed the APT experiments at different temperatures and demonstrated the influence of analysis temperature on D concentration profile14. Their studies also illustrated that even a small change in the local chemistry influences the D distribution substantially14,18. This work also exemplifies the impact of the analysis conditions on the quantitative deuterium distribution in materials, which was further evidenced in the study of bulk deuterides and hydrides22–25.

Walck & Hren1; and Kellogg & Panitz13 studied the trapping of implanted deuterium by structural defects such as vacancies induced by He implantation and a grain boundary in W. They found that deuterium, trapped by crystalline defects at cryogenic temperatures, diffuses and is subsequently released from the specimen as the temperature is raised. Hence, their studies1,13 strongly suggest the necessity of a cryogenic workflow according to which the charged specimen is immediately quenched in order to retain the hydrogen in the reversible trapping sites of the specimen and further analysis on it is carried out quickly2,28. This was further highlighted by more recent work reporting on hydrogen in steel16,20. These workflows involve the charging of APT tips with hydrogen or deuterium that are suitable for direct atom probe analysis2,19,28.

However, these workflows were developed for wire-type samples prepared by electrochemical polishing, which do not allow for site-specific analyses. Breen et al. demonstrated that a substantial amount of hydrogen was introduced during the preparation of specimens by electrochemical polishing, leading to many of the trapping sites being saturated and preventing effective deuterium charging20. Yet substantial progress in the field of APT has been achieved by the use of focused-ion beam (FIB) combined with scanning-electron microscopes (SEM) to prepare specimens from specific microstructural features29. As hydrogen interacts differently with different microstructural features, studying their trapping behaviour to guide alloy design principles requires developing systematic workflows for charging the site-specific APT lift-outs with hydrogen or deuterium.

The current study describes our journey through three hydrogen/deuterium charging routes and their associated workflows: cathodic charging, plasma charging and gas charging suitable for charging specimens prepared by site-specific lift-out, as opposed to more conventional electropolished wires. We chose a twinning induced plasticity (TWIP) steel as our model system for the current study since it is highly susceptible to the hydrogen embrittlement3,30. The role of stacking faults, Σ3 twin boundaries and random boundaries has also been discussed in the literature with respect to their contribution to the hydrogen embrittlement resistance of TWIP steels30–32. Yet, the actual prevalent hydrogen embrittlement mechanism is not well understood. We discuss the details of these workflows with pros and cons, in order to help the community avoid some of the pitfalls associated with hydrogen and deuterium charging of APT specimens.

Experimental details

Materials

High manganese TWIP steels contain over 20 wt.% Mn33 and are austenitic, with a face centred cubic (FCC) crystal structure. A model TWIP steel with a chemical composition of Fe 28Mn 0.3C (wt.%) was used for the current study. It was strip cast and subsequently homogenized at 1150°C for 2 hours. It was then 50% cold-rolled and recrystallized at 800°C for 20 minutes, followed by water cooling to room temperature.

Methods

A FEI Helios NanoLab 600i dual-beam FIB/SEM was used for preparing specimens for APT. APT experiments were conducted on either a LEAP 5000 XS or XR instrument (CAMECA Instruments Inc. Madison, WI, USA), in voltage pulsing mode at a set point temperature of 70K, 0.5% detection rate, 15–20% pulse fraction and 200kHz pulse repetition rate. These conditions had been defined in previous studies targeting an accurate detection of carbon in other austenitic steels34. APT data analyses were carried out using AP Suite 6.1 software.

For the cryogenic transfer workflows, we used the facilities developed within the framework of the Laplace project at the Max-Planck-Institute for Iron Research (MPIE), and detailed in ref. 35. This contains two atom probes, a CAMECA LEAP 5000 XS and a CAMECA LEAP 5000 XR, a FEI Helios Xe-plasma Focussed Ion Beam (PFIB), a N2-atmosphere glovebox (Sylatec) and a gas-charging chamber known as the Reacthub Module (RHM)36. These instruments are equipped with docking stations to host ultra-high vacuum carry transfer suitcases (UHVCTS, Ferrovac VSN40S) that are used for transferring the cryogenically cooled specimens between various instruments while minimizing contamination or frosting. The Xe-plasma FIB is also equipped with a cryogenically-cooled stage which allows cryo-FIB preparation of APT specimens and an intermediate chamber that enables the specimen transfers from the UHV suitcase to the cryo-stage. Cryogenic pucks were used to hold the APT specimens in the current study which are thermally insulated to avoid any direct contact with the vacuum transfer rods via a 2 mm-thick layer of polyether ether ketone (PEEK).

Workflows

Cathodic/electrolytic charging

Cathodic/electrolytic charging of metallic specimens is carried out in an electrolytic cell, consisting of an electrolyte, i.e., the charging solution, the sample as the cathode, a Pt wire as the anode and a reference electrode connected to a power source. The electrolytic cell used in the present study is shown in Figure 1 (a) and was developed and used previously for wire-type samples prepared by electrochemical polishing – see ref. 20. The process involves immersing the wire-specimen held inside the cryo-puck into the charging solution. Here, the aim was to develop the cathodic charging process for site-specific APT specimens mounted on a support. The conventional geometry used for APT site-specific specimens is a silicon coupon support held by a clip and mounted on a copper stub29,37. However, Cu would dissolve in the solution, hindering the use of this conventional geometry.

0d399c15-e179-4670-8b18-c31498a953e6_figure1.gif

Figure 1.

(a) Set up for cathodic charging (electrolytic cell); (b) charged specimen inserted into an Al puck holder block immersed in LN2 for an immediate quench.

Here, specimens were prepared on molybdenum grids, which are typically used for transmission electron microscopy (TEM) specimens and previously reported as a suitable alternative to support lift-outs for APT38,39. Figure 2 (a) shows an optical micrograph of the molybdenum grid with a diameter 3 mm which is cut using a razor blade into a half grid. The size of each Mo grid post is approx. 50 µm. The Mo grid is first electropolished in an aqueous solution of 10% NaOH (Sigma-Aldrich, Germany) by applying a DC voltage of ~ 10V38–40. After electropolishing, the size of Mo grid post is reduced to approx. 10 µm, as shown in the optical micrograph in Figure 2 (b). This is then further sharpened by a Ga-FIB using annular milling process29 in order to make the posts 2–3 µm in diameter, as visible from the SEM image in Figure 2 (c). The grid is held by a grid holder following the design detailed in ref. 38, and is used for mounting the samples for hydrogen charging. The grid holder is then inserted into the cryo-puck35, which was fixed to the holder connected to the cathode of the electrolytic cell, shown in Figure 1 (a).

0d399c15-e179-4670-8b18-c31498a953e6_figure2.gif

Figure 2.

(a) Molybdenum grid of diameter 3-mm; cut into (b) a half grid; (c) Mo grid post of 2–3 microns in diameter suitable for mounting the samples.

Site-specific APT specimen preparation was performed onto this Mo grid held in the grid holder following the protocol outlined in ref. 29. The geometry is illustrated in Figure 3, with a schematic in Figure 3 (a). Here, “t” refers to the thickness of the Pt weld applied for mounting the specimen on the Mo grid post. Figure 3 (b) shows the SEM image of side view of the Pt weld obtained by tilting the grid holder at an angle of 90° while inserting it into the SEM chamber. Figures 3 (c) and (d) are scanning electron micrographs showing the top view of the Pt weld when the grid alignment is horizontal and vertical, respectively.

0d399c15-e179-4670-8b18-c31498a953e6_figure3.gif

Figure 3.

(a) Schematic of the Pt weld where “t” is the thickness of Pt weld; (b) Scanning electron microscopy (SEM) image of the side view of Pt weld obtained by tilting the grid holder at 90° while inserting it into the SEM chamber and the thickness of Pt weld is 3 µm in this case; (c) SEM image of the top view of Pt weld when the grid alignment is horizontal and the thickness of Pt weld is 1 µm in this case; (d) SEM image of the top view of Pt weld when the grid alignment is vertical and the thickness of Pt weld is 3 µm in this case.

Figure 4 summarizes the seven workflows trialled, using three charging routes.

0d399c15-e179-4670-8b18-c31498a953e6_figure4.gif

Figure 4. Schematic depicting seven different workflows corresponding to three different charging routes: cathodic charging, plasma charging and gas charging.

APT = atom probe tomography; SS = stainless steel; TEM = transmission electron microscopy; PFIB = plasma focussed ion beam.

The detailed procedure of all workflows is described in the subsequent sections.

Workflow W1. Workflow W1 involves the preparation of a site-specific mounted lift-out using conventional Pt-welds, formed by decomposition of the Pt-precursor by using the Ga-ion beam, with a thickness of approx. 1 µm. The mounted lift-out is subjected to hydrogen charging prior to sharpening the deposited chunks into needle-shaped specimens. An aqueous solution of 0.05 M H2SO4 (Sigma-Aldrich, Germany, 98%) was used as an electrolyte for charging41. 1.4 g/l of thiourea (Merck Millipore, Germany) was mixed into the charging solution to act as hydrogen recombination binder41. It inhibits the combination of H atoms and prevents the formation of molecular H2, thereby enhancing the hydrogen ingress into the specimen. A DC voltage of 1.5V was applied.

After 1 hour of charging at room temperature, with the charging cell on a bench in air, the sample was immediately inserted into an Al puck holder block immersed in liquid nitrogen (LN2) for immediate quenching, as shown in Figure 1 (b). The Al block holding the specimen was transferred through ambient atmosphere into the cryogenically-cooled stage of the PFIB through an intermediate chamber that can be used as an air-lock, and not through the UHV suitcase. Upon pumping this intermediate chamber to high-vacuum conditions, the puck was inserted into the PFIB with the help of the dedicated transfer rod – see ref. 35 for details – in order to sharpen the specimen through annular milling29. Frost was observed on the Mo grid, as readily visible in Figure 5 (a). This could have been formed because the charging and transfers were performed in ambient atmosphere, which contains moisture. In this case, the sample was transferred back to the transfer rod of the PFIB35 and allowed to stay on it for 5 minutes in order to desorb the frost under high vacuum of 10-7 mbar28 and ambient temperature. The specimen was then transferred back to the cryostage for sharpening. Figure 5 (b) shows an SEM image of the specimen during the sharpening, which indicates that the Pt-welds were strongly affected by the charging or the frosting.

0d399c15-e179-4670-8b18-c31498a953e6_figure5.gif

Figure 5.

(a) Scanning electron microscopy (SEM) image of the frost observed on the Mo grid after it was transferred onto the plasma focussed ion beam (PFIB) cryostage following the H-charging in an aqueous solution of 0.05M H2SO4 and quenching in air; (b) SEM image of the specimen during sharpening in the PFIB.

Workflow W2. To avoid the formation of frost, all subsequent charging experiments were performed in a N2 filled glovebox (Sylatec). The specimen, electrolytic cell, charging solution and liquid nitrogen (LN2) bath with the Al puck holder were first loaded into the N2 glovebox. The LN2 bath inside the N2 glovebox must be filled for quenching the sample immediately after charging17. After filling the LN2 bath and precooling the Al block puck holder, the charging of the specimen was started in the electrolytic cell. After performing the charging and quenching in the N2 atmosphere, the samples were transferred into the PFIB through the precooled UHV suitcase35.

Specifically, in Workflow W2, we mounted the lifted-out chunk by using approx. 3 µm thick Pt-welds in the PFIB, formed by decomposition of the Pt-precursor using Xe-plasma FIB. The charging solution was the same as the one used in Workflow W1. Here again, the Pt-welds did not survive and their condition, after charging and quenching, was same as that of the previous experiment (Workflow W1), shown in Figure 5 (b).

Workflow W3. As a next step, in Workflow W3, since the Pt-welds appeared to be the weaker part of the mounted sample, we prepared a new batch of specimens by using e-beam induced Pt deposition to form larger and denser welds, with a thickness of approx. 3 µm. Subsequently, these specimens were charged following the same workflow as previously (see Workflow 2). SEM imaging of the specimens in the PFIB after transfer through the precooled UHV suitcase revealed that the Pt-weld had survived but the sample had corroded, as shown in Figure 6. This experiment evidences that the steel sample corrodes in the acidic charging solution, leading to the dissolution of the specimen during charging. The duration of charging is also a critical factor to be considered because a longer charging duration could also lead to the dissolution of the specimen, beyond the issue of the weak Pt-welds.

0d399c15-e179-4670-8b18-c31498a953e6_figure6.gif

Figure 6. Scanning electron microscopy image of the specimen after H-charging in an aqueous solution of 0.05M H2SO4 and transferring through the precooled ultra-high vacuum suitcase during sharpening in the plasma focussed ion beam.

Workflow W4. In the next experiment, Workflow W4, we decided to change the charging solution and use Ga-FIB induced Pt deposition to prepare specimens with 3 µm thick Pt-welds. A neutral solution consisting of an aqueous solution of 3 wt.% NaCl (Sigma-Aldrich, Germany) with 0.3 wt.% NH4SCN (Sigma-Aldrich, Germany, 98%) as hydrogen recombination binder was used for charging42. The neutral charging solution is comparatively less effective for charging because only H2O is the source of H+ ions in this solution. These H+ ions move towards the cathode and penetrate into the sample, which is connected to the cathode. Na+ and Cl- ions only enhance the ionic conductivity of the solution, whereas H2SO4 was also the source of H+ ions in the previously used acidic charging solution. Nevertheless, it should be sufficient for charging the APT specimens (Pt-welds) due to their extremely small, i.e., microscopic, surface area.

After charging the specimen in the neutral charging solution for 1 hour, the sample and weld both survived, the sample did not corrode and it was sharpened successfully in the PFIB. Figure 7 (a) is a SEM image of the sharpened specimen, which was taken to the LEAP 5000 XS for measurement through the precooled UHV suitcase. The corresponding mass spectrum is shown in Figure 7 (b), in which a substantial amount of Mo was observed. Mo could have originated from dissolution of the grid itself by the charging solution. At the applied voltage (1.5V) to the neutral electrolyte, Fe and Mo could have formed a galvanic couple due to which the following reaction can take place at the cathode:

0d399c15-e179-4670-8b18-c31498a953e6_figure7.gif

Figure 7.

(a) Scanning electron microscopy image of the tip sharpened in the plasma focussed ion beam cryostage after charging in the neutral solution and quenching; (b) the corresponding mass spectrum showing Mo peaks.

H2MoO4+6H++3e−↔Mo3++4H2OE°→+0.43V

Mo3+ from the solution would have deposited onto the sample, which was observed in the mass spectrum43.

Exchange of the charging solution. Both the acidic and the alkaline solutions, typically used for electropolishing of Mo grid, must be avoided for charging the steel of interest. Consequently, it was decided to use pure D2O (Sigma-Aldrich, Germany, 99.9 %) for charging and shorter charging times ranging from 1 to 5 minutes.

Some pure D2O charging experiments were also performed on non-site-specific lift-outs sharpened APT specimens. These needle-shaped APT specimens were prepared from the steel sample by electropolishing20 and then fine milled using the PFIB. The needle was charged in the N2 glovebox in pure D2O for 1 minute. The applied voltage was 2.2V, to induce the dissociation of the D2O electrolyte into D+ and OD- ions17. Then it was immediately quenched and transferred to the LEAP 5000 XS for measurement through the precooled UHV suitcase. Clusters of H2O, D2O and associated complex ions were found in the mass spectrum consistent with the analysis of water by APT44,45, but as soon as ions from the metal started to be detected, the specimen fractured.

For the next experiment, the sample was transferred to the PFIB immediately after charging and quenching in the N2 glovebox through the precooled UHV suitcase. SEM shows frost on the charged needle (see Figure 8 (a)), associated with the freezing of the remaining water adsorbed on the specimen’s surface. The specimen was therefore cleaned and resharpened in the PFIB post charging and quenching, Figure 8 (b), before being transferred to the LEAP 5000 XR for measurement.

0d399c15-e179-4670-8b18-c31498a953e6_figure8.gif

Figure 8.

(a) Scanning electron microscopy (SEM) image of the non-site specific lift-out specimen having the needle shaped geometry, surrounded by ice post charging and immediate quench; (b) SEM image of the same needle after being resharpened in the plasma focussed ion beam cryostage.

In a subsequent experiment, we tried to clean the charged needle with ethanol immediately before quenching. The sample was then transferred to the LEAP 5000 XS for measurement directly. However, this induced a slight delay in the quenching process – albeit only by few seconds. The mass spectrum exhibited peaks likely associated to adsorbed ethanol frozen on the specimen’s surface, and the specimen also fractured early.

Workflow W5. A set of pure D2O charging experiments was performed on a new batch of specimens prepared with approx. 3 µm thick Pt-welds formed by Ga-FIB induced Pt deposition. Multiple experiments were performed to optimize the charging time as charging for longer durations was leading to early specimen failure during measurement and charging for few seconds is not adequate to charge the sample with Pt-welds as the volume of material to charge is large compared to the needle-shaped geometry. Eventually, the charging performed for 5 minutes resulted in acceptable specimen yield. An SEM image of a sharpened specimen is shown in Figure 9 (a), and the corresponding 3D elemental map of carbon and deuterium are shown in Figures 9 (b) and (c), respectively, following reconstruction.

0d399c15-e179-4670-8b18-c31498a953e6_figure9.gif

Figure 9.

(a) Scanning electron microscopy image of the tip sharpened in the plasma focussed ion beam cryostage after charging the specimen in pure D2O and quenching; (b) 3D elemental map of carbon; and (c) deuterium, following its reconstruction.

This final workflow demonstrates successful electrolytic charging of site-specific APT specimens. The (re)sharpening of the specimen in the PFIB following charging can, however, cause the introduction of frozen D2O from the surface because of implantation by the incoming high-energy ions21. D2O molecules could also dissociate and lead to D ingress into the sample during milling. These effects can be detrimental to the analysistical performance but can be minimized or avoided by using lower acceleration voltages and lower beam current during (re)sharpening for instance.

Plasma charging (Workflow W6)

A plasma cleaner (Evactron), shown in Figure 10 (a), was used for plasma charging. The plasma cleaner is normally used for removing carbon contamination from SEM samples with an oxygen plasma. Here the plasma cleaner was fed by a gas line of Ar containing 5% H2 gas. For plasma charging, the samples were prepared conventionally on the silicon coupon held by a clip onto a copper stub, as the reactivity of copper in solution is no longer a concern.

0d399c15-e179-4670-8b18-c31498a953e6_figure10.gif

Figure 10.

(a) Plasma cleaner set up; (b) atom probe tomography specimen being charged with the hydrogen plasma; (c) the specimen placed on the precooled Al block for immediate quench post charging; (d) the specimen being transferred to the intermediate chamber of plasma focussed ion beam (PFIB); (e) scanning electron microscopy (SEM) image of the tip post charging and quenching; (f) SEM image of the same tip after resharpening in PFIB; (g) the corresponding mass spectrum.

Workflow W6 illustrates the plasma charging protocol. The APT specimens were prepared conventionally in the Ga FIB37 and were charged by the hydrogen-rich plasma for 1 h (shown in Figure 10 (b)) at a pressure of 10-1 mbar. Subsequently, they were immediately quenched in LN2 as shown by Figure 10 (c), and were transferred into the PFIB, on the pre-cooled cryostage through the intermediate chamber as depicted by Figure 10 (d). The SEM micrograph in Figure 10 (e) shows no frost on the specimen’s surface. It was resharpened at low kV (16 kV ion acceleration voltage and 30 pA beam current). Figure 10 (f) shows the SEM image of the tip after cleaning.

Finally, it was transferred into the LEAP 5000 XR for measurement through the precooled UHV suitcase and analysed. The corresponding mass spectrum is shown in Figure 10 (g) and eventually, the specimen fractured after only 100000 ions. During the APT analysis, hydrogen was detected in the mass spectrum at 1, 2 and 3 Da but its origin cannot be discerned. It could be from the specimen or from the analysis chamber of the atom probe. In addition, localised clusters of H2O, H3O were observed. These could originate from the vacuum in the plasma cleaner, which cannot be below 10-2 mbar, and hence not devoid of oxygen and moisture that introduced inside the specimen along with the hydrogen during charging. It might also be from mild frosting after plunge freezing and the introduction of O-containing surface species with the Xe-beam in the PFIB during sharpening.

Due to the relatively poor vacuum and associated implantation of impurities, we did not pursue this route for many more attempts. In principle, plasma cleaner directly attached to some commercial instruments could also be directly used to implant H by changing the gas supply for instance, which would enable these implantations in a cleaner environment and avoid the transfer through the UHV suitcase. However this has not, so far, been attempted.

Gas charging (Workflow W7)

Gas charging was carried out in the RHM36, as illustrated by Workflow W7. The RHM is equipped with a laser and hydrogen/deuterium gas flow so as to perform charging at a selected temperature while the sample is held on a cryo-stage to enable fast quenching. The samples are prepared on a cold-rolled stainless steel 304 (SS304) TEM half-grid, provided by JPT and CAMECA, as the RHM’s pyrometer is calibrated for this steel grade. Site-specific APT specimens were prepared by using the lift-out procedure described above with the Ga FIB37. It was transferred into the LEAP 5000 XR loadlock where it was subjected to an outgassing heat treatment at 150°C and 10-7 mbar pressure for approximately 4 hours in order to desorb the hydrogen previously trapped within the material, as suggested in ref. 20. This short heat treatment is not expected to substantially modify the microstructure but help desorb some of the trapped H within the specimen’s microstructure arising from the initial specimen preparation.

The specimen was subsequently transferred into the analysis chamber and an APT analysis was started in order to clean the specimen’s surface up to a voltage of 3–4 kV. The mass spectrum obtained from this pre-charging analysis shows only a peak at 1 Da, as shown in Figure 11 (a). It is expected that this hydrogen originates primarily from the residual gas of the analysis chamber, but could also in part be from the hydrogen trapped inside the material that was not desorbed by the outgassing heat treatment. Then it was transferred to the RHM for charging through the UHV suitcase. Deuterium gas charging was carried out in the RHM at a pressure of 250 mbar for 6 hours at 200°C, followed by an immediate quench down to 45K on the cryostage of the RHM. The similar deuterium gas charging conditions have been used for charging high strength steels15. The specimen was then transferred back into the LEAP 5000 XR for further measurement through the precooled UHV suitcase. The post charging mass spectrum exhibits peaks at 2, 3 and 4 in addition to the one at 1 Da as shown in Figure 11 (b); which confirms the D charging of the specimen.

0d399c15-e179-4670-8b18-c31498a953e6_figure11.gif

Figure 11.

(a) The pre-charging mass spectrum; (b) the mass spectrum after D gas charging in the Reacthub Module at a pressure of 250 mbar for 6 hours at 200°C.

In Figure 11 (b), the peak at 1 Da has higher counts in the mass spectrum following charging than that of the tip before charging (Figure 11 (a)). Further, Table 1 shows the bulk composition analysis in ionic %. H/Fe ratio is also higher in the post-charging mass spectrum than in the pre-charging mass spectrum. It implies that the residual background hydrogen is higher when the charged specimens are loaded into the analysis chamber. It could be misinterpreted as hydrogen originating from the specimen if the pre-charging mass spectrum had not been first collected and the intensity of the electric field is not compared between the pre- and post-charging measurements. The average charge state ratios of Fe+/Fe2+, corresponding to the pre- and post-charging mass spectra are comparable, as depicted in Table 1, indicating the comparable electric field in both analyses.

Table 1. Bulk composition analysis (ionic %) corresponding to the pre- and post-charging mass spectra.

Pre-charging
mass spectrum
Post-charging
mass spectrum
Amount of H (ionic %)1.436%3.193%
Amount of D (ionic %)00.1%
HFeratio0.0190.049
Fe+Fe2+ratio0.0090.011

Results and discussion

We have tried to explore several hydrogen/deuterium charging methods for charging the site-specific APT specimens. The current study is hence important to determine the hydrogen trapping behavior of specific microstructural features and their contribution to hydrogen embrittlement of a material. However, the hydrogen charging of APT tips and the accurate deuterium quantification by APT is challenging which are discussed in the present study.

Table 2 summarizes all the workflows corresponding to three different charging methods and their outcomes. The success of two workflows is evidenced in Table 2 - Workflow W5 through cathodic charging and Workflow W7 through gas charging.

Table 2. Summary of three different charging routes.

Charging
method
Charging parameters
WorkflowPt weld (‘t’: thickness)Charging solutionCharging
time
Voltage
applied (V)
Cryogenic
transfer
atmosphere
Result
Cathodic
charging
W1t - 1µm (Ga FIB)0.05M H2SO4 aq. soln +
1.4 g/l thiourea
1 hour1.5AirX
W2t - 3µm (PFIB)0.05M H2SO4 aq. soln +
1.4 g/l thiourea
1 hour1.5N2 Glovebox and
UHV suitcase
X
W3t - 3µm (e-beam)0.05M H2SO4 aq. soln +
1.4 g/l thiourea
1 hour1.5N2 Glovebox and
UHV suitcase
X
W4t - 3µm (Ga FIB)3 wt.% NaCl + 0.3 wt.%
NH4SCN aq. soln
1 hour1.5N2 Glovebox and
UHV suitcase
X
W5t - 3µm (Ga FIB)Pure D2O5 minutes2.2N2 Glovebox and
UHV suitcase
✓
Plasma
charging
W6Ga FIB sharpened APT tip
on Si coupon
Ar-5% H2 plasma (10-1
torr)
1 hourAirX
Gas
charging
W7Ga FIB sharpened APT tip
on 304 SS TEM half grid
250 mbar D gas6 hoursUHV suitcase✓

APT = atom probe tomography; FIB = focussed ion beam; PFIB = plasma-focussed ion beam; SS = stainless steel; TEM = transmission electron microscopy; UHV = ultra-high vacuum.

In cathodic charging routes (Workflows W1-W5), the mounted lift-out specimens were subjected to the cathodic hydrogen charging because a sharpened APT tip would have dissolved in the charging solution. A mounted lift-out prepared with 1 µm Pt-weld was also strongly affected by the charging, as observed in Workflow W1. The size of Pt-weld was hence optimized in the subsequent experiments and thick Pt-welds of 3 µm were prepared to enable the successful charging in the solution. The charged mounted lift-out specimens further needed to be sharpened by FIB milling for APT analysis.

In Workflow W5, the mounted lift-out chunk with approx. 3 µm thick Pt-weld prepared by Ga-FIB induced Pt deposition was cathodically charged in pure D2O in a N2 glovebox for 5 minutes at an applied voltage of 2.2 V. It was then immediately quenched in a LN2 bath and transferred to the PFIB cryostage for sharpening through the precooled UHV suitcase. The sharpened specimen was transferred to the atom probe for measurement through the precooled UHV suitcase. Workflow W7 consists of deuterium gas charging of a sharpened APT tip in RHM at a pressure of 250 mbar for 6 hours at 200°C followed by an immediate quench down to 45 K on the cryostage of the RHM.

Gas charging enables data acquisition from the same specimen before and after charging, which is useful for distinguishing between hydrogen coming from charging and hydrogen coming from the analysis chamber. This is not possible in the cathodic charging route, as the mounted lift-out chunk is charged before it is sharpened to enable APT analysis, so only post-charging data can be acquired. The pre-charging mass spectrum is essential for determining the amplitude of the background hydrogen signal, as had been pointed out by Walck and Hren1. They also collected several mass spectra from a Ni specimen prior to D-implantation. Also, no other specimen must be allowed to enter into the analysis chamber after the pre-run so as to avoid any change in the vacuum conditions of the analysis chamber due to the hydrogen coming from loading of other samples.

However, since the detection of H as atomic or molecular ionic species depends on the strength of the electrostatic field27, the relative H/D amounts across datasets can only be compared if they are acquired under similar electric field conditions. We estimated the electric field by calculating the average charge state ratios (CSR). We selected the peaks of C and CSR is computed as 12C2+/12C1+ by using the peaks at m/q = 6 Da and m/q = 12 Da, containing 12C2+ and 12C1+, respectively. Admittedly the field evaporation behaviour can be more complex46 and can be subject to additional detection issues47,48; the carbon peaks were selected because they can be found across all datasets, conversely to Fe1+ which was sometimes not detected above background. Multiple datasets of charged and uncharged specimens were used from both the LEAP 5000 XS and LEAP 5000 XR atom probes for plotting the average charge state ratios vs. relative H abundances in Figure 12 (a) and (b), respectively.

0d399c15-e179-4670-8b18-c31498a953e6_figure12.gif

Figure 12.

(a) Average charge state ratios (CSR) vs. relative H abundances for the experiments performed in LEAP 5000 XS; (b) the average CSR vs. relative H abundances for the experiments performed in LEAP 5000 XR; (c) enlarged region delineated by the black box in Figure 12 (b).

Three relative H abundances are plotted in each: H1/Htotal, H2/Htotal and H3/Htotal where Htotal is (H1 + H2 + H3). H1, H2 and H3 correspond to peaks obtained at 1, 2 and 3 Da, respectively. The precision in the measurement of all the involved species is computed as their counting statistics12:

σi=Ci×(1−CiN

where σi is the precision (counting statistics), Ci is the atomic fraction of the element i, N is the number of measured ions which lie in between 100000 to 30 х 106 ions for the measurements used for plotting these curves.

The x-axis error bar corresponding to the CSR is computed as:

σCSR=(σC2+)2+(σC1+)2

The y-axis error bar corresponding to the three relative H abundances is computed as:

σH1/Htotal,σH2/HtotalandσH3/Htotal=(σH1)2+(σH2)2+(σH3)2

All reported data from the LEAP 5000 XS are from cathodically charged specimens with hydrogen, except one H-gas charged in the RHM at room temperature at a pressure of 250 mbar. All reported from the LEAP 5000 XR are from D-gas charged specimens in the RHM except the one which was plasma charged (highlighted in light green).

Figure 12 (c) is a close-up on a section of charge-state ratios. The region highlighted in red qualitatively illustrates the region with higher relative H2 and H3 contents than the region delineated by the red line that contains mostly data from uncharged specimens. The amount detected remains relatively low but appears significantly higher than the background and above the level in the uncharged specimens. It has been reported that higher loading fugacities are achieved using cathodic charging than plasma charging, which is higher than low pressure gas charging2,49. However, gas charging using the RHM is the cleanest approach for charging APT specimens prepared by site-specific lift-out. The moisture content encountered in the mass spectra of specimens subjected to gas charging was negligible and the transfer of specimens to the PFIB for cleaning was also not required after gas charging.

We could later explore the possibility of equipping the RHM with a plasma generator for instance, or make use of the one directly on the commercial LEAPs, which would help maintain a cleaner UHV-chain. This would enable more facile analyses of plasma-charged specimens, through a cleaner approach compared to electrolytic charging, with quicker ingress kinetics than gas charging while allowing pre- and post- charging data acquisition. However, radiation damage from the incoming energetic ions may also cause additional modification of the microstructure that should not be disregarded.

Conclusion

We have described three hydrogen/deuterium charging routes for charging site-specific APT specimens in the present study: cathodic charging, plasma charging and gas charging. A detailed step by step description of the corresponding workflows involved in optimizing the three charging routes is also given. The merits and drawbacks of the three charging methods are illustrated in the present study, while we demonstrate the success of a cathodic charging workflow and the gas charging approach. Although higher ingress kinetics may be achieved by plasma charging, it may also damage and introduce defects into the microstructure. The origin of hydrogen/deuterium can still be questioned in the case of cathodic charging due to its possible introduction during (re)sharpening of the specimen in the PFIB. Hence, we infer that the gas charging in the RHM provides the cleanest amongst our results, i.e. with least contamination, and would be the selected route for quantification of hydrogen/deuterium in the studied TWIP steel.

An earlier version of this article can be found on arXiv (arXiv:2109.13650v1).

Data availability

Underlying data

Zenodo: Hydrogen and deuterium charging of site-specific specimen for atom probe tomography. https://doi.org/10.5281/zenodo.553485950.

There are 28 pos files in the above provided Zenodo link which contain raw data corresponding to all APT datasets used for the current study and one range file in .RRNG format to index all these data files.

Data are available under the terms of the Creative Commons Attribution 4.0 International license (CC-BY 4.0).

Acknowledgements

The authors would like to thank Uwe Tezins, Andreas Strum and Christian Bross for their support to the FIB, APT, UHV suitcase, Reacthub Module and N2 glovebox facilities at MPIE. Ms. Monika Nellessen and Ms. Katja Angenendt are acknowledged for their support in SEMs and metallography sample preparation labs. H.K., S-H.K., L.T.S. and B.G. acknowledge the financial support from the ERC-CoG-SHINE-771602. R.S.V acknowledges financial support from the IMPRS SurMat scholarship.

References

  • 1.  Walck S, Hren J: FIM / IAP / TEM STUDIES OF HYDROGEN IN METALS To cite this version: HAL Id: jpa-00224446 of 1984. J Phys Colloques. 1984; 45(C9): C9–360. Publisher Full Text
  • 2.  Haley D, Merzlikin SV, Choi P, et al.: Atom probe tomography observation of hydrogen in high-Mn steel and silver charged via an electrolytic route. Int J Hydrogen Energy. 2014; 39: 12221–9. Publisher Full Text
  • 3.  Koyama M, Akiyama E, Lee YK, et al.: Overview of hydrogen embrittlement in high-Mn steels. Int J Hydrogen Energy. 2017; 42: 12706–23. Publisher Full Text
  • 4.  Wan D, Deng Y, Barnoush A: Hydrogen embrittlement effect observed by in-situ hydrogen plasma charging on a ferritic alloy. Scr Mater. 2018; 151: 24–7. Publisher Full Text
  • 5.  Bhadeshia H: Prevention of Hydrogen Embrittlement in Steels. ISIJ International. 2016; 56(1): 24–36. Publisher Full Text
  • 6.  Barrera O, Bombac D, Chen Y, et al.: Understanding and mitigating hydrogen embrittlement of steels: a review of experimental, modelling and design progress from atomistic to continuum. J Mater Sci. 2018; 53(9): 6251–90. PubMed Abstract | Publisher Full Text | Free Full Text
  • 7.  Dwivedi SK, Vishwakarma M: Hydrogen embrittlement in different materials: A review. Int J Hydrogen Energy. 2018; 43: 21603–16. Publisher Full Text
  • 8.  Sun B, Lu W, Gault B, et al.: Chemical heterogeneity enhances hydrogen resistance in high-strength steels. Nat Mater. 2021; 2021: 1–6. PubMed Abstract | Publisher Full Text
  • 9.  Merzlikin SV, Borodin S, Vogel D, et al.: Ultra high vacuum high precision low background setup with temperature control for thermal desorption mass spectroscopy (TDA-MS) of hydrogen in metals. Talanta. 2015; 136: 108–13. PubMed Abstract | Publisher Full Text
  • 10.  Luo H, Li Z, Chen YH, et al.: Hydrogen effects on microstructural evolution and passive film characteristics of a duplex stainless steel. Electrochem Commun. 2017; 79: 28–32. Publisher Full Text
  • 11.  De Geuser F, Gault B: Metrology of small particles and solute clusters by atom probe tomography. Acta Mater. 2020; 188: 406–15. Publisher Full Text
  • 12.  Gault B, Chiaramonti A, Cojocaru-Mirédin O, et al.: Atom Probe Tomography. Nat Rev Methods Prim. 2021; 1–51. Publisher Full Text
  • 13.  Kellogg GL, Panitz JKG: A direct observation of the trapping of deuterium ions at a grain boundary in tungsten. Appl Phys Lett. 1980; 37: 625. Publisher Full Text
  • 14.  Gemma R, Al-Kassab T, Kirchheim R, et al.: Analysis of deuterium in V-Fe5at.% film by atom probe tomography (APT). J Alloys Compd. 2011; 509: S872–6. Publisher Full Text
  • 15.  Takahashi J, Kawakami K, Kobayashi Y: Origin of hydrogen trapping site in vanadium carbide precipitation strengthening steel. Acta Mater. 2018; 153: 193–204. Publisher Full Text
  • 16.  Chen YS, Haley D, Gerstl SSA, et al.: Direct observation of individual hydrogen atoms at trapping sites in a ferritic steel. Science. 2017; 355(6330): 1196–9. PubMed Abstract | Publisher Full Text
  • 17.  Chen YS, Lu H, Liang J, et al.: Observation of hydrogen trapping at dislocations, grain boundaries, and precipitates. Science. 2020; 367(6474): 171–5. PubMed Abstract | Publisher Full Text
  • 18.  Gemma R, Al-Kassab T, Kirchheim R, et al.: Visualization of deuterium dead layer by atom probe tomography. Scr Mater. 2012; 67: 903–6. Publisher Full Text
  • 19.  Gemma R, Al-Kassab T, Kirchheim R, et al.: APT analyses of deuterium-loaded Fe/V multi-layered films. Ultramicroscopy. 2009; 109(5): 631–6. PubMed Abstract | Publisher Full Text
  • 20.  Breen AJ, Stephenson LT, Sun B, et al.: Solute hydrogen and deuterium observed at the near atomic scale in high-strength steel. Acta Mater. 2020; 188: 108–20. Publisher Full Text
  • 21.  Chang Y, Lu W, Guénolé J, et al.: Ti and its alloys as examples of cryogenic focused ion beam milling of environmentally-sensitive materials. Nat Commun. 2019; 10(1): 942. Publisher Full Text
  • 22.  Sundell G, Thuvander M, Andrén HO: Hydrogen analysis in APT: methods to control adsorption and dissociation of H₂. Ultramicroscopy. 2013; 132: 285–9. PubMed Abstract | Publisher Full Text
  • 23.  Kolli RP: Controlling residual hydrogen gas in mass spectra during pulsed laser atom probe tomography. Adv Struct Chem Imaging. 2017; 3(1): 10. PubMed Abstract | Publisher Full Text | Free Full Text
  • 24.  Chang YH, Mouton I, Stephenson L, et al.: Quantification of solute deuterium in titanium deuteride by atom probe tomography with both laser pulsing and high-voltage pulsing: Influence of the surface electric field. New J Phys. 2019; 21: 053025. Publisher Full Text
  • 25.  Mouton I, Breen AJ, Wang S, et al.: Quantification challenges for atom probe tomography of hydrogen and deuterium in Zircaloy-4. Microsc Microanal. 2018; 25(2): 481–8. PubMed Abstract | Publisher Full Text
  • 26.  McCarroll IE, Bagot PAJ, Devaraj A, et al.: New frontiers in atom probe tomography: a review of research enabled by cryo and/or vacuum transfer systems. Mater Today Adv. 2020; 7: 100090. PubMed Abstract | Publisher Full Text | Free Full Text
  • 27.  Tsong TT, Kinkus TJ, Ai CF: Field induced and surface catalyzed formation of novel ions: A pulsed-laser time-of-flight atom-probe study. J Chem Phys. 1983; 78(7): 4763–75. Publisher Full Text
  • 28.  Chen YS, Bagot PAJ, Moody MP, et al.: Observing hydrogen in steel using cryogenic atom probe tomography: A simplified approach. Int J Hydrogen Energy. 2019; 44(60): 32280–91. Publisher Full Text
  • 29.  Prosa TJ, Larson DJ: Modern Focused-Ion-Beam-Based Site-Specific Specimen Preparation for Atom Probe Tomography. Microsc Microanal. 2017; 23(2): 194–209. PubMed Abstract | Publisher Full Text
  • 30.  An D, Krieger W, Zaefferer S: Unravelling the effect of hydrogen on microstructure evolution under low-cycle fatigue in a high-manganese austenitic TWIP steel. Int J Plast. 2020; 126(2): 102625. Publisher Full Text
  • 31.  Kwon YJ, Seo HJ, Kim JN, et al.: Effect of grain boundary engineering on hydrogen embrittlement in Fe-Mn-C TWIP steel at various strain rates. Corros Sci. 2018; 142: 213–21. Publisher Full Text
  • 32.  Kwon YJ, Jung SP, Lee BJ, et al.: Grain boundary engineering approach to improve hydrogen embrittlement resistance in Fesingle bondMnsingle bondC TWIP steel. Int J Hydrogen Energy. 2018; 43(21): 10129–40. Publisher Full Text
  • 33.  Gutierrez-Urrutia I, Raabe D: Dislocation and twin substructure evolution during strain hardening of an Fe-22 wt.% Mn-0.6 wt.% C TWIP steel observed by electron channeling contrast imaging. Acta Mater. 2011; 59(16): 6449–62. Publisher Full Text
  • 34.  Yao MJ, Dey P, Seol JB, et al.: Combined atom probe tomography and density functional theory investigation of the Al off-stoichiometry of κ-carbides in an austenitic Fe–Mn–Al–C low density steel. Acta Mater. 2016; 106: 229–38. Publisher Full Text
  • 35.  Stephenson LT, Szczepaniak A, Mouton I, et al.: The Laplace project: an integrated suite for correlative atom probe tomography and electron microscopy under cryogenic and UHV conditions. PLoS One. 2018; 13(12): e0209211. PubMed Abstract | Publisher Full Text | Free Full Text
  • 36.  Khanchandani H, El-Zoka AA, Kim SH, et al.: Laser-equipped gas reaction chamber for probing environmentally sensitive materials at near atomic scale. 2021. Reference Source
  • 37.  Thompson K, Lawrence D, Larson DJ, et al.: In situ site-specific specimen preparation for atom probe tomography. Ultramicroscopy. 2007; 107(2–3): 131–9. PubMed Abstract | Publisher Full Text
  • 38.  Herbig M, Choi P, Raabe D: Combining structural and chemical information at the nanometer scale by correlative transmission electron microscopy and atom probe tomography. Ultramicroscopy. 2015; 153: 32–9. PubMed Abstract | Publisher Full Text
  • 39.  Felfer PJ, Alam T, Ringer SP, et al.: A reproducible method for damage-free site-specific preparation of atom probe tips from interfaces. Microsc Res Tech. 2012; 75(4): 484–91. PubMed Abstract | Publisher Full Text
  • 40.  Makineni SK, Lenz M, Kontis P, et al.: Correlative Microscopy-Novel Methods and Their Applications to Explore 3D Chemistry and Structure of Nanoscale Lattice Defects: A Case Study in Superalloys. JOM. 2018; 70: 1736–43. Publisher Full Text
  • 41.  Zan N, Ding H, Guo X, et al.: Effects of grain size on hydrogen embrittlement in a Fe-22Mn-0.6C TWIP steel. Int J Hydrogen Energy. 2015; 40(33): 10687–96. Publisher Full Text
  • 42.  Sun B, Krieger W, Rohwerder M, et al.: Dependence of hydrogen embrittlement mechanisms on microstructure-driven hydrogen distribution in medium Mn steels. Acta Mater. 2020; 183: 313–28. Publisher Full Text
  • 43.  Miller MK: Atom Probe Tomography: Analysis at the Atomic Level. 2000. Publisher Full Text
  • 44.  Schwarz TM, Weikum EM, Meng K, et al.: Field evaporation and atom probe tomography of pure water tips. Sci Rep. 2020; 10(1): 20271. PubMed Abstract | Publisher Full Text | Free Full Text
  • 45.  El-Zoka AA, Kim SH, Deville S, et al.: Enabling near-atomic-scale analysis of frozen water. Sci Adv. 2020; 6(49): eabd6324. PubMed Abstract | Publisher Full Text | Free Full Text
  • 46.  Peng Z, Zanuttini D, Gervais B, et al.: Unraveling the Metastability of Cn2+ (n= 2–4) Clusters. J Phys Chem Lett. 2019; 10(3): 581–588. PubMed Abstract | Publisher Full Text | Free Full Text
  • 47.  Sha W, Chang L, Smith GDW, et al.: Some aspects of atom-probe analysis of Fe-C and Fe-N systems. Surf Sci. 1992; 266(1–3): 416–23. Publisher Full Text
  • 48.  Peng Z, Vurpillot F, Choi PP, et al.: On the detection of multiple events in atom probe tomography. Ultramicroscopy. 2018; 189: 54–60. PubMed Abstract | Publisher Full Text
  • 49.  Farrell K, Lewis M: The Hydrogen Content of Austenite After cathodic charging. Scripta Metallurgica. 1981; 15(6): 661–664. Publisher Full Text
  • 50.  Khanchandani H, Kim SH, Varanasi RS, et al.: Hydrogen and deuterium charging of site-specific specimen for atom probe tomography. 2021. http://www.doi.org/10.5281/zenodo.5534859