Introduction

In 2020, 2.7 million people in the European Union were diagnosed with cancer, and another 1.3 million people died as consequence of the same disease. Today, Europe hosts a tenth of world’s population but contributes to a quarter of world’s cancer cases. Lives lost to cancer are expected to rise by more than 24% by 2035. Although cancer is an individual diagnosis, it has non-negligible impacts on the lives of the patient’s relatives and social group. The overall economic impact of cancer in Europe is estimated to cost more than €100 billion per year1.

In addition to traditional cancer treatments such as surgery, chemotherapy, immunotherapy and classical radiation therapy, during the last 60–70 years hadron therapy emerged as a new field in radiation therapy. In particular it demonstrated to be a key option for treatments of pediatric solid tumours, tumours of the brain, skull base, spinal cord, upper respiratory tract, chest, pelvic and others2.

CERN is active in knowledge and technology transfer from physics to medicine and society in general. In 1996 the Proton-Ion Medical Machine Study (PIMMS)3,4 group was formed with commitment to design a cancer therapy synchrotron. CERN agreed to host this study in its Proton Synchrotron Division. One of the many results of this study has been the implementation of a synchrotron and a fixed gantry in the medical environment of the Italian National Center of Oncological hadron therapy (CNAO) and MedAustron in Austria.

With the definition of the European Strategy for Particle Physics in 2006, in one of its statements, CERN Council recognized the need to promote the impact of particle physics research on society proposing ‘to create a technology transfer forum to analyse the keys to the success in technology transfer projects in general’5. Since 2006 regular updates have been required by the European Strategy Group. Environmental and societal impact has been one of the pillars in every update, the most recent being in 20206. Meanwhile, to further formalize the commitment of CERN in knowledge and technology transfer in medical applications, in 2014 the CERN Medical Applications Steering Group was founded by council decision, being this the very first organizational structure created with this scope. In 2016 a more robust structure was established7 guided by CERN Medical Applications Steering Commitee.

As CERN did with PIMMS in the past, its commitment to develop next-generation ion therapy facilities resulted in the establishment of the Next Ion Medical Machine Study (NIMMS)8, an initiative funded by the CERN Knowledge Transfer for Medical Applications and intended to stimulate wide European collaborations. NIMMS is organised in three work packages (superconducting magnets, linacs, gantries). The aim of the gantries work package is to design a superconducting ion gantry that is more efficient in term of costs and size with respect to existing ones, both normal-conducting and superconducting. Indeed, ion therapy requires a rotating gantry to maximize its efficacy giving to clinicians a tool that is able to provide 3D optimized treatments, necessary for example to better spare an organ at risk or to minimize the radiation dose in healthy surrounding tissues. Existing gantries for heavy ions, are massive (300 tons for the superconducting unit of HIMAC in Chiba9 and 600 tons for the normal-conducting one of HIT in Heidelberg10) and require high standard technologies, thus expensive. The cost makes these facilities a luxurious investment for countries, limiting general public accessibility to such treatments.

In the development of gantry technologies two projects approved and funded by European Commission Horizon 2020 Research Infrastructure are integrating the NIMMS resources with those of collaborating institutes throughout Europe, namely: HITRI plus (Heavy Ion Therapy Research Integration)11 for the development of ion therapy accelerator technologies, and I.FAST (Innovation Fostering in Accelerator Science and Technology) for the development of curved superconducting magnets. These activities are additionally supported by the State Research Programme in High Energy Physics and Accelerator Technologies of Latvia12. The combined efforts of these projects aim to answer the needs of the medical accelerator community for latest state of the art technologies associated to ion therapy.

Main developments on rotating isocentric gantries before the start of HITRI plus in April 2021 were concentrated on an innovative scenario named SIGRUM (Superconducting Ion Gantry with Riboni’s Unconventional Mechanics13,14), developed by TERA Foundation in collaboration with CERN. A following study was mainly focused on a possible mechanical optimization of the structure15.

This note reports on a new collaborative effort between CERN EN-MME (Mechanical and Materials Engineering department), CERN TE-MSC (Magnets, Superconductors and Cryostats Technology department), RTU (Riga Technical University) and UNIBS (University of Brescia) consulting experts from the CNAO ion therapy centre as part of the HITRI plus project. Several scenarios of mechanical structures for isocentric gantries have been identified, a series of parameters to evaluate the compliance to main requirements for the selected scenarios have been defined and eventually the result of the comparison is reported.

Challenges

The first major challenge is the control of the cost of the complete and functional machine that has to remain within some price limits defined in the HITRI plus project16. Since a reduction in the load carried by the mechanical structure is expected from the use of bent superconducting magnets of new design, there are opportunities to reduce weight and cost. Moreover, a second challenge is to provide a comprehensive and impartial analysis to obtain reliable results while still in a preliminary phase of the project. The multidisciplinary nature of the gantry is in itself another relevant challenge, requiring to acknowledge interdisciplinary connections and to evaluate their influence on studied parameters. In general terms, the improvement of the Technology Readiness Level (TRL)17 of this design is the final outcome of the project. Starting from the presently evaluated TRL 1 for the mechanical structure and its motion system, whose technology concepts have been proposed, reaching TRL 2.

Main requirements

Table 1 summarizes the main desirable characteristics for the next generation of rotating gantries.

Table 1. List of relevant desirable characteristics for a rotating gantry.

Safetyrobustness of the design
Technological prerequisitessize
weight and complexity
performances: deformation and precision
Environmentalminimal energy consumption
Clinical requirementstreatment angle coverage

Design robustness is guaranteed by relevant mechanical design choices. This requirement obviously comes as first rule while designing a machine or component that meets safety specifications.

Size requirements were set in accordance with CNAO based on expected available on-site space for a possible first implementation of the gantry, corresponding to an internal volume of about 20×20×20m3 room (width × length × height). The HITRI plus project has placed specific requirements on weight and complexity to favour a subsequent industrialization of the design. The target in this case is to maintain the weight at about 100 tons, which is a substantial reduction when compared to the 600 for HIT (with a gantry hall of 15×26×16m3) and 300 tons of HIMAC respectively (gantry length of 13m and beam orbit radius of 5.45m, estimated gantry hall of 15×20×16m3). Deformation and precision performances requirements are obtained by translating the needs of the medical community to ensure the proper accuracy in the beam delivery to the patient.

Environmental requirements, mainly related to the minimization of energy consumption, shall be considered as well. The machine consumes energy in multiple systems: cryogenic, transfer line magnets, accelerator magnets and gantry motion. However, the consumption of the latter is orders of magnitude lower than the others; its relevance is proportionally weighted in our study, since it is not heavily impacting the total efficiency of the system.

Important clinical requirements come from the necessity of maximizing effectiveness of treatment plans by using multiple treatment angles, thus a gantry rotation between ±110° and ±180° (measured from the horizontal plane) is required.

Methods

The study has been done using a comprehensive approach aiming at being impartial in providing objective tools to compare mechanics advantages and disadvantages of each proposed scenario. This note reports about challenges related to the mechanical structure and its motion system, without considering aspects related to medical treatment. However, general inputs from clinicians have been considered, such as: estimated size of patient couch and its motion, desired volumetric imaging system at the isocenter and the gantry angular speed. The study was subdivided into three steps:

This study has evolved directly from the SIGRUM analysis13,14. The proposal of scenarios was carried out during periodic meetings. Comparison parameters were proposed by taking into account the ones related to project requirements and challenges, with the expansion or addition of new parameters. The analysis of such parameters has been done by using the following tools:

Results or hypotheses have been validated by double checking with different methods in the case of quantitative data (i.e. for mathematical models) or, if qualitative, by reaching common agreement during constant discussions within the group of experts mentioned above. Results have been summarized in a final table by ranking the scenarios with respect to each parameter. The arbitrary ranking system consists in levels from one to five, related to a rising level of compliance to the specific parameter. During the study development, the awareness of crucial medical needs for appropriate treatment was always present.

Comparison and analysis of scenarios for rotating gantry

The transfer line proposed in 13,14 was used as general input for this study:

The comparison of scenarios was done assuming independence of the mechanical structures from the transfer line general layout unless its mass and radial position of the center of gravity varies much (i.e. using 5T magnets). However, even in that case it is believed that relative advantages and disadvantages would not change much between scenarios. Between 10 and 15 scenarios, considering both balanced and unbalanced variants were proposed during discussion sessions with experts. However, not every scenario satisfied all general technological requirements (i.e. design robustness, energy consumption or size limits). Therefore, a screening process has been applied, mainly driven by considerations related to the robustness of the design and failure scenarios:

Table 2. Prospective scenarios of mechanical structure for rotating isocentric gantries.

bb09c384-3678-4d2f-b088-b751c43b3879_T2a.gifBalanced SIGRUM (B-SIGRUM)

A balanced structure with two side supports, driven by an electric motor (and gearbox) directly
mounted on the axis. Proposed to improve robustness of the design of SIGRUM only by the addition of
a relatively simple counterweight (in green).
•  ±110° deliverable treatment angle range.
•  External volume of construction: 21.5 × 28.5 × 27m3
•  Internal volume of construction: 11.5 × 20.5 × 20m3
•  Rotating mass about 70 ton.
•  Relevant moment of inertia: 500000 kgm2
bb09c384-3678-4d2f-b088-b751c43b3879_T2b.gifSimply Supported Gantry (SSG)

Two side supports and one perimetrical support balanced structure. Driven by an external electric
motor through off-axis rope and pulley transmission system. Proposed to reduce absolute
displacement without unconventional compensation methods.
•  ±110° deliverable treatment angle range.
•  External volume of construction: 22×28.5× 30m3
•  Internal volume of construction: 12×20.5× 23m3
•  Rotating mass about 70 ton.
•  Relevant moment of inertia: 500000 kgm2
bb09c384-3678-4d2f-b088-b751c43b3879_T2c.gifSide and Cradle Supported Gantry (SCSG)

One side and one ground support balanced structure. Driven by off-axis friction rollers, coupled to
an electric motor, or by external hydraulic system through rope transmission system. Proposed as
intermediate solution between SIGRUM and Full Turn Gantry.

•  ±110° deliverable treatment angle range.
•  External volume of construction: 21.5 × 28.5 × 27m3
•  Internal volume of construction: 11.5 × 20.5 × 20m3
•  Rotating mass about 120 ton.
•  Relevant moment of inertia: 700000 kgm2
bb09c384-3678-4d2f-b088-b751c43b3879_T2d.gifFull Turn Gantry (FTG)

Two ground support balanced structure, driven by off-axis friction rollers, coupled to electric motor.
Proposed to cover a full turn, minimizing bed couch movements.
•  ±180° deliverable treatment angle range.
•  External volume of construction: 28×28.5×27m3
•  Internal volume of construction: 16×20.5×20m3
•  Rotating mass about 120 ton.
•  Relevant moment of inertia: 700000 kgm2

Prospective scenarios

The scenarios that sufficiently satisfied requirements mentioned in the previous paragraph were more thoroughly investigated. They are the balanced variants shown in Table 2, their main properties being listed in Table 3.

Table 3. List of scenarios main properties.

B-SIGRUMSSGSCSGFTG
Treatment angle range±110±110±110±180deg
External volume of construction21.5 × 28.5 × 2722 × 28.5 × 3021.5 × 28.5 × 2728 × 28.5 × 27m3
Internal volume of construction11.5 × 20.5 × 2012 × 20.5 × 2311.5 × 20.5 × 2016 × 20.5 × 20m3
Rotating mass7070120120ton
Relevant moment of inertia500000500000700000700000kgm2

The size of the support ring varies between SCSG and FTG, influenced by a different positioning of the patient couch which has to compensate the limited rotation of ±110° of the SCSG.

Dimensions of 20 × 20 × 20 m3 (width × length × height) have been specified by the responsible physicists of a medical facility that is interested in the installation of the machine; the study of the present article is intended to be as general as possible and new facilities may have less stringent requirements. Relating the size requirement to the cost of the facility, cost which is highly influenced by the volume of the building, all the proposed structures have an overall internal volume lower than 20 × 20 × 20 m3. All structures exceed the length by 0.5 meters, which is not impacting the comparison. Only SSG structure exceeds the height by 3 meters, while keeping the overall internal volume less than the required one. Additionally, the height is considered to be of less concern in the space reservation, not impacting the allocated footprint.

Analysis of main parameters

Taking into account the technological requirements in Table 1 and the challenge related to costs, Table 4 lists a series of decisive parameters that have been proposed and analysed following the methodology explained in the homonym Section. The minimum clinical requirement of ±110° rotation is satisfied by all scenarios proposed, while only the FTG can cover the full ±180° rotation.

Table 4. List of key parameters.

Robustness of the design
•  Failure scenariosConsider the possibility of the gantry to accelerate uncontrollably due to gravity
or a failure of the motion system
•  Possible safety brakes positionEvaluates the possibility to have safety brakes in strategical positions
•  Number of systems to brakeEstimates the number of systems to brake
Size
•  Room space requirementsEvaluates the volume of the scenario
Weight and complexity
•  Weight and inertiaEvaluates the weight and moment of inertia of scenarios
•  ComplexityEvaluates the complexity of the assembly i.e. R&D needed for the supports
Performances: deformation and precision
•  Load propertiesExpress the max value of the load on the supports and if it is related to the
angular position of the gantry during the treatment phase.
•  Deformation performances (structure)Evaluates the scenarios based upon displacements of the transfer line
•  Improvement margin of deformation performancesEstimates the margin of mechanical optimization based upon knowledge
•  Deformation performances (supports)Assesses the deformation of the supports
•  TolerancesEstimates the difficulty in achieving tolerances in different scenarios
Costs
•  Cost of manufacturing componentsEstimates relative costs of manufacturing components for each scenario
•  Cost of the driving systemEstimates relative costs of the driving system
Environmental
•  Minimal energy consumptionEstimates the energy consuption
•  Material activationEstimates the impact of material activation for the gantry structure

The main decisive parameters are explained in the following sections, those not described explicitly are thought to be deductible from the summary Table 6 that will appear at the end of the analysis of all parameters.

Failure scenarios

Addressing failure scenarios, two key aspects have been considered: the acceleration due to gravity, and the malfunctioning of the driving system. The model of non-damped physical pendulum has been taken into account with Equation 1 in the case of an unbalanced gantry accelerating due to gravity:

ϑ″=mgRCOGJsin⁡ϑ(1)

where: m is the mass of the gantry, g is the gravitational acceleration, RCOG is the radial position of the center of gravity with respect to the rotation axis, ϑ is the angular position of the gantry (measured from the top vertical position) and J is the moment of inertia of the gantry. Equation 2 models the case of a malfunction of the driving system for balanced scenarios.

ϑ″=τ/J(2)

where τ is the torque applied to the gantry. The torque τ is assumed to be constantly applied and be twice the maximum nominal value of the torque.

The software Mathematica has been used to calculate the rotational kinetic energy for a base case (SIGRUM) and for the other balanced scenarios proposed (Python could have been used alternatively to obtain same results). This was done by solving numerically the differential Equation 1 and Equation 2 calculating the angular speed with a starting value of 0.21 rad s−1 (as discussed above): results are reported in Figure 1. The value of 0.2 s represents the time before a collision of balanced structure with walls, assuming a clearance of 5° starting from the nearest available treatment position. This requirement shall also be used to design the treatment room environment avoiding dangerous collisions. Furthermore, the values of the kinetic energy after the reaction time have been used to rank structures in relation to the ‘failure scenarios’ parameter in Table 6.

bb09c384-3678-4d2f-b088-b751c43b3879_figure1.gif

Figure 1. Kinetic energy of proposed scenarios within a reaction time of 0.2s.

The values at t = 0.2s were used to rank the scenarios with the values reported on the figure.

Deformation of the structure and its improvement margin

Deformation performances of the structure were already addressed in the preliminary stage of the study: an estimation of displacement of transfer line elements was done for the SIGRUM in 15,18. Whenever a requirement needs to be satisfied from the mechanical point of view it must be noted that stricter requirements (i.e. lower displacement) usually require higher stiffness, adding weight to the structure. Therefore, to achieve the compromise between stiffness and weight within the requirements set for this study, an innovative method of mechanical optimization to recover the rigid part of the displacement was proposed15,18, reducing the displacements by one order of magnitude (from several mm to less than a mm, see Figure 2 and Figure 3). Aforementioned results have been obtained assuming to use stainless steel (Young's modulus 200 GPa) for the main structure due to magnetic properties.

bb09c384-3678-4d2f-b088-b751c43b3879_figure2.gif

Figure 2. View of the mechanical optimized structure of the SIGRUM, numbers refer to analysed nodes for the displacement recovery.

bb09c384-3678-4d2f-b088-b751c43b3879_figure3.gif

Figure 3. Out of plane (screen plane) displacement of analysed nodes (Figure 2) in function of their radial position with respect to the rotation axis.

The remaining displacement after recovery of its rigid component is the distance between nodes and the straight line through the origin.

This optimization process has been proved to be particularly effective if applied to structures that behave like ‘cantilever’ beams (with a portion of structure hanging while one end is fixed). There is no reason to believe that this method is not applicable to B-SIGRUM with almost equal results. Furthermore, even if applicable this method has not yet been applied neither to SCSG nor to FTG, because the iterative mechanical optimization has not been performed. However, it is possible to assume that similar results can be reached for SCSG and FTG. Although the aforementioned method is not applicable to the SSG scenario, this is expected to benefit of similar displacement reduction thanks exclusively to its support configuration, acting like a simply supported beam (a structure supported at both ends, see Table 5). The advantage between a simply supported structure and a cantilevered one is independent from the material properties, given that the two structure are made of the same material.

Table 5. Comparison of maximum deflection between cantilever and simply supported beams.

Where w is the weight per length unit, L is the length of the beam, E is the Young’s modulus of the material of the beam and I is the area moment of inertia of the cross section of the beam.

Supports configurationMaximum displacement
Cantilever beambb09c384-3678-4d2f-b088-b751c43b3879_T5a.gifηmax⁡=18wL4EI
Simply supported beambb09c384-3678-4d2f-b088-b751c43b3879_T5b.gifηmax⁡=5384wL4EI≈0.104×18wL4EI
bb09c384-3678-4d2f-b088-b751c43b3879_figure4.gif

Figure 4. Work done by the motor (ideal) for a 180° rotation divided in equal δ angle steps.

Table 6. Summary table of scenario rankings with respect to each decisive parameter listed in Table 4.

Robustness of the designFailure scenariosPossible safety brakes positionNumber of systems to brakeSizeRoom space requirementsWeight and complexityComplexityWeight and inertiaPerformances: deformation and precisionTolerancesDeformation performances (structure)Improvement margin of structure deformationDeformation performaces (supports)Load propertiesCostsCost of manufacturing componentsCost of the driving systemEnvironmentalEnergy consumption and material activation
SIGRUM53555555334415
SSG45533555534455
SCSG34554345443244
FTG34514345455244

All the scenarios have been ranked at the same level according to deformation performances (Table 6). An estimated improvement margin of structure deformation has been defined. This evaluates the possible improvement in deformation analysis deriving from the mechanical optimization of a scenario with respect to the current knowledge. Therefore, SSG, SCSG and FTG are ranked better than B-SIGRUM, in relation to this parameter (Table 6).

Deformation of supports

Each scenario has a different combination of weight and support configuration, therefore side supports of B-SIGRUM are mainly subject to bending moments while bottom supports of FTG are loaded by compression forces. Since for balanced scenarios, during the treatment phase, these loads are not related to the angular position of the gantry, the supports will deform and the axis will be displaced by a constant value: this systematic error can be compensated finding a new origin for the patient couch or using adjustment systems (i.e. rails and shims between the supports and the gantry) to fine tune the position of the gantry in the assembly phase. As the loads are different for each scenario, side supports have been designed in SOLIDWORKS and studied with preliminary FEA in order to reach the same performances, namely the same horizontal and vertical displacement. Thus, differences in load values are decoupled from the evaluation of deformation performances and incorporated in the cost analysis (cost of components manufacturing): to achieve the same displacement under different loads, different amounts of material are needed. However, due to the different type of loads, adjustment systems have respectively different design complexity. Therefore, scenarios that have side supports have a lower rate than FTG which has two ground supports, as represented in Table 6.

Geometrical tolerances

The errors due to tolerances can be compensated at different levels: between the gantry and its supports, between the gantry and cryostats, or fine tuning the position of magnets inside cryostats. Furthermore, the ability to reach strict tolerances is usually in competition with the size of the component to manufacture, therefore, it is believed that a relatively small shaft like for B-SIGRUM and SSG would be easier to manufacture within a given range of tolerance if compared with large rings of SCSG and FTG. It is underlined that tolerances were qualitatively discussed.

Costs of components manufacturing

Costs of manufacturing components were estimated on the basis of previous knowledge on costs of main manufacturing processes that are necessary for this machine. The total cost of manufacturing components is composed by:

Cmanufacturing=Crot−structure+Csupports+Ccounterweight(3)

where:

  • Crot-structure is the cost of the rotating structure manufactured in stainless steel.

  • Csupports is the cost of support structures manufactured in construction steel.

  • Ccounterweight is the cost of the counterweight manufactured in lead.

The rotating structure and supports necessary for the functioning of each scenario were modeled and checked in simplified FEA to estimate the amount of material necessary to satisfy requirements. For example, for side supports, the amount of material was estimated to satisfy a maximum displacement of L/500-L/250 (common values for steel structures in engineering sound practice, where L is the span) and to have the same displacement between side supports of different scenarios (computed to be around 5mm or L/1800). The supports of magnets cryostats and instrumentation are related to the given transfer line layout (same for all scenarios), thus their cost can be neglected in the comparison. Therefore, Table 6 reports a ranking associated to relative costs between scenarios: it has been estimated that B-SIGRUM and SSG are lower in cost by a factor of two with respect to SCSG and FTG. Given the high cost of other components of the machine, such as superconducting dipole magnets, cryogenic cooling system, civil engineering, the choice of material was judged of much less impact on the total cost; for instance, the total cost of the material for the structure in the worst case would be 1/3 of the estimated cost of a single superconducting dipole. In fact, the cost of raw material is estimated 0.35% to 1% of the total cost: the choice was then focused on a material that guarantees the best corrosion resistance and is commonly considered to fit the tight requirements for a medical machine. The use of the same material for the studied structures does not impact the relative comparison performed in this section for different structures.

Costs of the driving system

The costs of the driving system are related to the maximum power (25–40kW), torque and speed required by the load. These data were calculated by assuming the equation of motion of the gantry (the same for all scenarios) and as a function of the properties of the gantry (mass, inertia) and the geometric properties of the transmission (i.e for the FTG, the radius of the ring and rollers on which the gantry is supported). Scenarios are ranked mainly based on the output torque of the required gearbox. These costs are between one and two orders of magnitude smaller with respect to the cost of components manufacturing.

Minimal energy consumption

The evaluation of the energy consumption was done analytically by calculating the work of the motor for a 180° rotation subdivided in equal δ steps. The results are reported in Figure 4: the percentages shown on the plot represent the energy ratios between different scenarios and the SIGRUM, the base example. The energy ratios are evaluated for two reference values, δ = 10° and 45°, respectively related to the minimum angle step between approved treatment angles, and a possible realistic number of treatment angles for a patient (4–5 angles). The expected energy ratio will range between these two values. In conclusion, the motion system of B-SIGRUM and SSG consume less energy with respect to SCSG and FTG, all four consuming less than the SIGRUM.

Material activation

Detailed studies on material activation have not been conducted at the stage of the comparative analysis. However, further discussions have been done with experts in the field who operate a vertical beam of carbon ions: the proposed structures in Table 2 have been judged to not behave drastically differently from an activation point of view. This is because for the different scenarios the morphology of the structure in proximity of the transfer line is comparable with a similar amount of mass, leading to a similar relevant material activation.

In a conservative way, if it is assumed for sake of simplicity that activation is proportional to volume of the gantry structure, the solution B-SIGRUM and SSG behave better than solution SCSG and FTG. Scenarios are rated accordingly in Table 6, combining the energy consumption criteria with the material activation one. Following that bigger structure have bigger volume and higher inertia, those that consume less energy are also those that would have less material activated.

Results of the comparative study

All proposed scenarios are safe, however, SSG is estimated to be the safest. B-SIGRUM and SSG are expected to be lower in cost by a factor of two with respect to SCSG and FTG. Statically balanced scenarios, by means of a relatively simple integrated counterweight, come with several advantages. Although some of the following considerations were not estimated quantitatively, they are not negligible:

Conclusions

Multiple scenarios were defined during discussions with experts of CERN EN-MME group, TE-MSC group, RTU and UNIBS in consultancy with CNAO experts in the framework of NIMMS and HITRI plus project. Main requirements (Table 1) and challenges were acknowledged following HITRI plus16 requirements and general designing rules for a machine of this kind. The comparative study has been performed in a comprehensive manner following a unified methodology providing objective results, summarized in Table 6. The main conclusions are:

Data availability

Zenodo: Comparative Study on Scenarios for Rotating Gantry Mechanical Structures, https://doi.org/10.5281/zenodo.651684619. Data are available under the terms of the Creative Commons Attribution 4.0 International license (CC-BY 4.0).

References

  • 1.  European comission: Europe’s beating cancer plan. 2021; Last visited on 09/03/2022. Reference Source
  • 2.  CNAO: Pathologies that can be treated. Last visited on 22/02/2022. Reference Source
  • 3.  Badano L, Benedikt M, Bryant PJ, et al.: Proton-Ion Medical Machine Study (PIMMS), 1. Technical Report CERN-PS-99-010-DI, CERN-TERA Foundation-MedAustron Oncology-2000 Collaboration, 3. 1999. Reference Source
  • 4.  Bryant PJ, Badano L, Benedikt M, et al.: Proton-Ion Medical Machine Study (PIMMS), 2. Technical Report CERN-PS-2000-007-DR Accelerator Complex Study Group, 8. 2000. Reference Source
  • 5.  The European Strategy for particle physics (Brochure), 7. 138th Session of Special Restricted Council. 2006; Last visited on 10/03/2022. Reference Source
  • 6.  2020 Update of the European Strategy for Particle Physics (Brochure). 199th Session of CERN Council. CERN-ESU-015. 2020; Last visited on 10/03/2022. Reference Source
  • 7.  Strategy and framework applicable to knowledge transfer by CERN for the benefit of medical applications, 6. 304th Meeting of Scientific Policy Committee. CERN/SPC/1091; CERN/FC/6125; CERN/3311. 2017; Last visited on 10/03/2022. Reference Source
  • 8.  Next ion medical machine study. knowledge transfer, cern accelerating science. Reference Source
  • 9.  Iwata Y, Noda K, Shirai T, et al.: Design of a superconducting rotating gantry for heavy-ion therapy. Phys Rev Spec Top-Ac. 2012; 15(4): 044701. Publisher Full Text
  • 10.  Fuchs R, Weinrich U, Sust E: Assembly of the carbon beam gantry at the heidelberg ion therapy (hit) accelerator. Proceedings of the EPAC-2008. Genoa, Italy. 2008; 23–27. Reference Source
  • 11.  Heavy ion therapy research integration plus. 2021; Last visited on 10/03/2021. Reference Source
  • 12.  The state research programme in high energy physics and accelerator technologies of latvia. Last visited on 10/03/2022. Reference Source
  • 13.  Amaldi U, Alharbi N, Benedetto E, et al.: SIGRUM - A Superconducting Ion Gantry with Riboni’s Unconventional Mechanics. Technical Report CERN-ACC-NOTE-2021-0014; NIMMS-Note-002, 6. 2021; Last visited on 22/02/2022. Reference Source
  • 14.  Benedetto E, Al Harbi N, Brouwer L, et al.: A carbon-ion superconducting gantry and a synchrotron based on canted cosine theta magnets. 2021; arXiv preprint arXiv: 2105.04205. Publisher Full Text
  • 15.  Perini D, Dassa L, Piacentini L, et al.: Preliminary design of the support structure for a rotating carbon-ion transfer line for medical applications. Instruments. 2021; 5(4): 34. Publisher Full Text
  • 16.  Heavy ion therapy research integration. Reference Source
  • 17.  Technology readiness level, horizon 2020 – work programme 2014-2015. Reference Source
  • 18.  Piacentini L: Project development of a rotating transferring line for carbon ions used for medical scope. Master’s Thesis, unpublished, 2021.
  • 19.  Piacentini L, Dassa L, Perini D, et al.: Comparative Study on Scenarios for Rotating Gantry Mechanical Structures. 2022. http://www.doi.org/10.5281/zenodo.6516846